1、Model 38DL PLUSOperation Training,IndexClick on the links below to go directly to the corresponding slide.(Please Note:you have to be in slide show mode for the links to work.),Power Up,The On/Off key is used to turn the 38DL PLUS on and off.,Back toIndex,Model 38DL PLUSDual Element Do-ZERO,Back toI
2、ndex,Do-ZERO for Standard Dual Transducers,Press,Uncouple the transducer and make sure the tip is free of any couplant layer.,“Do-ZERO”allows the gage to recognize the transducer for optimal setup.It will also measures the time of flight through the transducer to compensate for transducer wear and c
3、hanges in temperature.,Note:The 2nd F,Zero can be pressed any time the transducer is uncoupled to update the Zero offset.,Back toIndex,then,2nd F,Do-ZERO for THRU-COAT Transducers,Uncouple the transducer and make sure the tip is free of any couplant layer.,Note:The Zero can be pressed at any time th
4、e transducer is uncoupled to update the Zero offset or to turn THRU-COAT On or Off.,“Do-ZERO”allows the gage to recognize the transducer for optimal setup.It will also measures the time of flight through the transducer to compensate for transducer wear and changes in temperature.,Press:,Use,to turn
5、THRU-COAT On or Off,Press EnterPress Enter on“OK”to complete the Do-ZERO,Back toIndex,then,2nd F,Selecting Measurement Units and Resolution,Use,to highlight the Measurement setup then press ENTER,Allows the user to set Measurement Units and Resolution,Press:,Use,to select“units”or“resolution”and,to
6、change the setting,then press Meas,Back toIndex,Main Screen,Back toIndex,Model 38DL PLUSCalibration Standard Dual Transducers,Back toIndex,Standard Dual Element Cal Velocity,Couple transducer to the thick sample press:,Allows the user to calibrate for the speed of sound of the material to be tested.
7、Usually done on a sample representing the maximum of the measurement range.,Uncouple the transducer and enter the known thickness,Once reading is steady press:,Back toIndex,ENTER,Standard Dual Element Cal Zero,A subtracted time measurement is used to compensate for the transit time through the delay
8、 line in the transducer and the couplant layer.Cal Zero is usually done on a sample representing the minimum of your measurement range.,Once reading is steady press:,Uncouple transducer and enter the known thickness.Then press the MEAS key to complete the calibration.,Couple transducer to the thin s
9、ample press:,Back toIndex,ENTER,Model 38DL PLUSTHRU-COAT Calibration,Back toIndex,THRU-COAT Cal Velocity,Couple transducer to the thick sample press:,Allows the user to calibrate for the speed of sound of the material to be tested.Usually done on a sample representing the maximum of the measurement
10、range.,Uncouple the transducer and enter the known thickness,Once the reading is steady press:,Back toIndex,ENTER,THRU-COAT Cal Zero,A subtracted time measurement used to compensate for the transit time through the delay line in the transducer and the couplant layer.Cal Zero is usually done on a sam
11、ple representing the minimum of your measurement range.,Once reading is steady press:,Uncouple transducer and enter the known thickness.Press the Meas key to complete the calibration or press the CAL VEL key to calibrate for the velocity of the coating.,Couple transducer to the thin sample press:,Ba
12、ck toIndex,ENTER,Coating Calibration,Couple transducer to a sample with known coating thickness press:,Allows the user to calibrate for the speed of sound of the of the coating material.It is not necessary to calibrate for the coating if the coating value is not being displayed or if the accuracy of
13、 the coating thickness is of less importance.,Uncouple the transducer and enter the known coating thicknessPress MEAS to complete the calibration,Once reading is steady press:,Back toIndex,ENTER,Directly enter the velocity in THRU-COAT Mode,Allows the user to directly enter the velocity of the mater
14、ial to be tested.When using THRU-COAT the velocity of the coating can also be entered.,The current velocity will bedisplayed.Use the arrow keysto enter the know velocity.Thenpress MEAS,Press,Press,The current coating velocity will bedisplayed.Use the arrow keysto enter the know velocity.Thenpress ME
15、AS,again,Back toIndex,2nd F,then,2nd F,then,THRU-COAT Measurement Screen,Back toIndex,Model 38DL PLUSCalibration of Single Element Transducers,Back toIndex,Single Element Transducer/Setup Selection,The 38DL PLUS has 21 preset default single element transducer setups stored in its permanent memory an
16、d up to 35 user defined custom setups can be entered.To select a default transducer setup press,The user can select one of the default transducer setups or any of the user defined custom setups.,Back toIndex,Then use the,keys to select from the available lists and press ENTER,Then use the,keys to se
17、lect a transducer setup and press the MEAS key to recall the setup and return to the measure mode.,Single Element Cal Velocity,Couple transducer to the thick sample and press:,Allows the user to calibrate for the speed of sound of the material to be tested.Usually done on a sample representing the m
18、aximum of the measurement range.,Uncouple the transducer and enter the known thickness use in the,keys,Once reading is steady press:,Back toIndex,ENTER,Single Element Cal Zero,A subtracted time measurement used to compensate for the transit time through the delay line in the transducer and the coupl
19、ant layer.Cal Zero is usually done on a sample representing the minimum of your measurement range.,Once reading is steady press:,Couple transducer to the thin sample press:,Back toIndex,ENTER,Uncouple the transducer and enter the known thickness use in the,keys,Model 38DL PLUSInstrument Lock with Pr
20、ogrammable Password,Back toIndex,Instrument Lock with Programmable Password,The instrument lock on the 38DL PLUS allows the user to lock advanced features and functions so that they are not accidently altered or changed.A password can be set so that the locked functions can not be unlocked without k
21、nowing the passwordThe following functions can be lockedCalibrationAccess to Setup Menu and SP MenuTransducer RecallDatalogger(except for the save key)The ability to adjust the gainWaveform adjustment parameters,Back toIndex,Note:If the user forgets the password please contact Olympus for the master
22、 password.,Instrument Lock,setting a Password,Press,Setting a password is an optional feature and the 38DL PLUS can be locked with out a password.,Back toIndex,Note:If the user forgets the password please contact Olympus for the master password.,Then use the,keys to select Password Set and press ENT
23、ER,Use the editing functions to enter a password and press Enter use the,keys to select set and press ENTER.,Instrument Lock with without Password,Press,If a password has been set then the password must be entered in the Instrument Lock screen before any of the functions can be used.If a password ha
24、s not been set then the password area will be grayed out.,Back toIndex,Then use the,keys to selectInstrument Lock and press ENTER,Use the editing functions to enter a password and press Enter use,keys to select the a function to lock and,lock and unlock it.Highlight set and press ENTER to activate t
25、he lock,Example of a locked function,Back toIndex,Model 38DL PLUSUsing the EMAT Transducer,Back toIndex,EMAT Transducer,An EMAT(Electro Magnetic Acoustic Transducer)uses the Magnetostrictive principle to generate shear wave sound energy in ferrous metals that are externally coated with high temperat